Gemnet IV gives The application log file is full message
Symptom
Gemnet IV gives The application log file is full message
Error message
The application log file is full Select Event Viewer from the Start \ programs \ administrative tools menu. Then from the Event Viewer menu select Log from menu bar. Select Settings from the drop down menu. Select Over Write files as needed for both Application and System logs . This will stop the event log full messages.