The Customer complains of a Water or a fluid leak on the system.
Symptom
The Customer complains of a Water or a fluid leak on the system. The customer indicates there is water leaking from the closed loop cooling system. Fluid is coming out of the tube arm assembly or in the technical room. If zero stock on 2184070-2 RP ROTATING JOINT then order 2129844-2 RIGID WATER COOLING ASSEMBLY and it contain the joint part. Also replace the two O-rings with new or they will leak next. The Customer complains of a Water or a fluid leak on the system. The customer indicates there is water leaking from the closed loop cooling system. Fluid is coming out of the tube arm assembly or in the technical room. 2184070-2 RP ROTATING JOINT then order 2129844-2 RIGID WATER COOLING ASSEMBLY and it contain the joint part. Also replace the two O-rings with new or they will leak next
Error message
error Code = 1200109, Class = IH, error Text = ERR_CONDITIONER_TANK_EMPTY, debug Text = [DemSupervisor::_bootConditioner] Water tank is empty~~, recovery Class = IH, major Function /= XRImDet System impact: X-ray unavailable. Displayed message : "Xray not available - Reset system; if problem persists, call Service" error Code = 1200109, Class = IH, error Text = ERR_CONDITIONER_TANK_EMPTY, debug Text = [DemSupervisor::_bootConditioner] Water tank is empty~~, recovery Class = IH , major Function /= XRImDet System impact: X-ray unavailable. Displayed message : "Xray not available - Reset system; if problem persists, call Service"
Applies to
- Models
- Innova
- Product
- Innova 2000 Innova 2000 systems Innova 2000 systems Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
- Component
- RIGID WATER COOLING ASSEMBLY in c-arc 2129844-2 RIGID WATER COOLING ASSEMBLY and it contain the joint part. Also replace the two O-rings with new or they will leak next. When cables pulls on the water connections at the Center of the C-arc cable hose the copper tube or the o-ring inside the connecter is bad.. Description: RIGID WATER COOLING ASSEMBLY FLS - RIGID WATER COOLING ASSEMBLY ADDED PER SPCN 280302JLV165852N1. USED ON LCV+ AND LC+. 2129844-2 REPLACES 2121687-2 PER SAME SPCN. MSA8429 04/15/02
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.