Lift failure, dose is minimized, call service.

Symptom

Lift failure, dose is minimized, call service. Detector lift will not move up or down. Even by using the emergency backout button. Lift failure, dose is minimized, call service. Detector lift will not move up or down. It also will not move with the emergency backout button. Lift failure, dose is not minimized, call service. Detector lift will not move up or down. It also will not move with the emergency backout button.

Error message

error Code = 1050254, error Text = Detector Lift Motor Failure, debug Text += Lift motor failure (4A77), recovery Class = WR, major Function /= GantryAxis System impact: Lift lock-up. Displayed message : "Lift failure, dose is not minimized, call service asap".

error Code = 50111, hw-status=0003 FRT-Lift error Text = POS SMA: Encoder, potentiometer failure, debug Text = pos-error=4ABC, hw-status=0003, usr-status=2000 , recovery Class = DO, major Function /= LC Gantry System impact: X-ray aborted & Gantry lock-up. Displayed message : None. 0x4ABC 0x0000 HW_FRT_L 0x0001 HW_FRT_PVT 0x0002 HW_FRT_ARC 0x0003 HW_FRT_II-Lift 0x0000 HW_FRT_L 0x0001 HW_FRT_PVT 0x0002 HW_FRT_ARC 0x0003 HW_FRT_Iift System impact: X-ray aborted & Gantry lock-up. Displayed message : None.

error Code = 50213, error Text = Detector lift failure, debug Text += 4A77, recovery Class = WR, major Function /= XRImChainPos System impact: Lift lock-up. Displayed message : "Lift failure, dose is not minimized, call service asap".

Applies to

Models
Innova
Product
Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova 2000 Innova 2000 systems Innova 2000 systems Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
Component
Positioner Gantry Positioner Gantry Positioner Gantry