System only sees the Head Coil ID code, no matter what coil is connected.
Symptom
System only sees the Head Coil ID code, no matter what coil is connected. Coil ID Fails. See Long Description The customer was scanning with the head coil, then finished that patient and connected the CTL coil and started a new exam using the USCTLTOP coil. But the system wouldn't permit it, it posted a coil ID error that stated that it saw a coil code that I later found was the same one as the head coil, and it was expecting to see the coil ID code for the CTL coil. So somehow the Head coil status was retained by or "stuck" in the coilID hardware/software. The IH2 engineer arrived and verified the CTL was plugged in. She shut the system down and cycled complete system power and rebooted and that fixed the problem. It seems the SRI2 must have been at fault here. I suspect the SRI2 somehow retained the code from the previous coil and rebooting and/or just cycling the power obviously reset that state. I think the problem could've been fixed by just cycling SSM power (to reset the SRI2) and then TPS Reset but she hadn't tried that. The IH2 engineer will run SRI2 diagnostics soon. Thu Aug 17 07:43:34 2006 Error: 2220007 Host: rmnq Process: scn File: RxMgrDownload.m Line: 2138 The prescribed RF coil and detected RF coil do not match. Upper coil connect status:1, Upper coil code: bv523rifp0t9r4h3pgqceh64jh7d2n2d , Lower coil connect status:0, Lower coil code:Null, Rx coil name:USCTLTOP, Rx coil code:pS4ONH8K1VniMi8PAhceHrFjZQ1txgaU ----------------------------------------------------------------------------------------------- You can see that the coil ID code seen by the system is the same as the head coil. Also it shows Upper coil connect status:1, Lower coil connect status:0, but there isn't anything plugged into the upper port. the CTL coil is the only coil plugged in, and it's plugged into the lower port. [insite@rmnq config]$ more CoilConfig.cfg comment.com = "This is the Excite-II 1.5T MR CoilConfig file." coilFileRev = "11.0.M3a" coil1 = { coilName = "HEAD" \ coilCode = " bv523rifp0t9r4h3pgqceh64jh7d2n2d " \ physicalCoilDescription = "Standard Head Coil by GE" \ coil36 = { coilName = "USCTLTOP" \ coilCode = "pS4ONH8K1VniMi8PAhceHrFjZQ1txgaU" \ physicalCoilDescription = "4 Ch CTL Array by USAI" \ -------------------------------------------------------------------- This is the /usr/g/service/log/review.out file that shows the prescription that they had downloaded when the error occurred. date: 8/17/2006 time: 7:43:34 ID : 06-345-702 PATIENT INFORMATION SCREEN Weight (kg) :65.8 Rx Level :EXAM Rx Mode :ACTIVE Scanner Mode :RESEARCH PATIENT POSITION SCREEN Entry :HEAD Position :SUPINE Coil Name: USCTLTOP Landmark :406.6000 Protocol :C-Spine for Myelopathy/1 Series De:upper scout IMAGING PARAMETERS SCREEN Image Mode :2D Scan Plane:SAGITTAL Pulse Seq :SPIN ECHO Grad Mode :WHOLE Image Options :None Psd Name :memp SATURATION PARAMETER SCREEN SAT Pulses :NO SAT PULSE!! Supress Tech :NONE Tag Type :None Tag Space :7 IMAGE ENHANCE SCREEN Surface Coil Intensity Correction :No Filter Choice :None USER CV SCREEN SCAN TIMING SCREEN Number of Echoes :1 Echo Time (TE) :20.0 Rep Time (TR) :250.0 RBW :15.6 kHz SCANNING RANGE SCREEN Field of View :48 cm Slice Thickness :5.0 mm ISpacing:2.5 mm Start Loc :L25.0 P-0.0 I-0.0 End Loc :R27.5 P-0.0 I-0.0 # Slices :8 ACQUISITION TIMING SCREEN Acq. Matrix :256 X 128 NEX :1.00 Freq. Dir :S/I Phase FOV :1.00 Contrast :No Auto CF :WATER Prescan Options:AutoShim GENERAL INFORMATION Est. Ave. SAR :0.9415 Est. Peak SAR :1.8829 dB/dt Limit : 66% Scan Time :0:38 # Acquisitions:1 Table Delta :0.0 Research Mode -------------------------------------------------------------------- Software running on the system: [insite@rmnq ~/bin]$ /usr/g/bin/getver Hostname : rmnq Build number for cclass is 11.0_M4_0403.a Build number for driverSupport is 11.0_M4_0403.a Build number for insite is 11.0_M4_0403.a Build number for install is 11.0_M4_0403.a Build number for MrpApps is 11.0_M4_0403.a Build number for mx_apps is 11.0_M4_0403.a Build number for mx_plat is 11.0_M4_0403.a Build number for mx_terra is 11.0_M4_0403.a Build number for OS is GEMS_Linux_1.8.9 Build number for patch_11.0_M4_00 is 11.0_M4_0403.a.PA Build number for patch_11.0_M4_01 is 11.0_M4_0403.a.PQ Build number for patch_11.0_M4_02 is 11.0_M4_0403.a.PI Build number for patch_11.0_M4_03 is 11.0_M4_0403.a.PA Build number for patch_11.0_M4_05 is 11.0_M4_0403.a.PH Build number for patch_11.0_M4_08 is 11.0_M4_0403.a.PB Build number for patch_11.0_M4_09 is 11.0_M4_0403.a.PR Build number for patch_11.0_M4_10 is 11.0_M4_0403.a.PP Build number for patch_11.0_M4_11 is 11.0_M4_0403.a.PP Build number for patch_11.0_M4_12 is 11.0_M4_0403.a.PP Build number for patch_11.0_M4_13 is 11.0_M4_0403.a.PL Build number for patch_11.0_M4_14 is 11.0_M4_0403.a.PL Build number for patch_11.0_M4_15 is 11.0_M4_0403.a.PF Build number for patch_11.0_M4_16 is 11.0_M4_0403.a.PD Build number for patch_11.0_M4_17 is 11.0_M4_0403.a.PG Build number for patch_11.0_M4_18 is 11.0_M4_0403.a.PL Build number for patch_11.0_M4_21 is 11.0_M4_0403.a.PM Build number for patch_11.0_M4_22 is 11.0_M4_0403.a.PF Build number for patch_11.0_M4_23 is 11.0_M4_0403.a.PF Build number for patch_11.0_M4_24 is 11.0_M4_0403.a.PK Build number for patch_11.0_M4_25 is 11.0_M4_0403.a.PJ Build number for patch_11.0_M4_26 is 11.0_M4_0403.a.PH Build number for patch_11.0_M4_27 is 11.0_M4_0403.a.PI Build number for patch_11.0_M4_28 is 11.0_M4_0403.a.PI Build number for patch_11.0_M4_29 is 11.0_M4_0403.a.PI Build number for patch_11.0_M4_30 is 11.0_M4_0403.a.PR Build number for patch_11.0_M4_31 is 11.0_M4_0403.a.PI Build number for patch_11.0_M4_32 is 11.0_M4_0403.a.PN Build number for patch_11.0_M4_33 is 11.0_M4_0403.a.PL Build number for patch_11.0_M4_34 is 11.0_M4_0403.a.PQ Build number for patch_11.0_M4_36 is 11.0_M4_0403.a.PR Build number for patch_11.0_M4_37 is 11.0_M4_0403.a.PM Build number for patch_11.0_M4_38 is 11.0_M4_0403.a.PM Build number for patch_11.0_M4_44 is 11.0_M4_0403.a.PS Build number for pc is 11.0_M4_0403.a Build number for PostSdC is 11.0_M4_0403.a Build number for TERRA is IOS_M4_20031119_R292a_M4_Patch MR Software release: 11.0_0403a
Error message
2218630 The download of a prescribed series failed. 2218630 Host: Andys3T Process: scn File: RxManagerIF.m Line: 556 The download of a prescribed series failed. Internal status code is 2218501.
2220007 The prescribed RF coil and detected RF coil do not match. See long description on reading/interpreting this type of error message Note: An example of this error message, with explanation, is provided below. Wed Mar 3 08:18:58 2004 Error: 2220007 Host: lx-mr3 Process: scn File: RxMgrDownload.m Line: 2074 The prescribed RF coil and detected RF coil do not match. 040303104912225 Upper coil connect status:0, Upper coil code:Null, Lower coil connect status:1, Lower coil code:q8dCUW0PD2EaETIplAp90PLrACM6CW38, Rx coil name:MRI 8HRKNRR PA, Rx coil code:q8dCUWOPD2EaETIpIAp9oPLrACM6CW38 NOTE: Upper coil connector status refers to connector plugged into upper port at "dog house" (i.e. head coil or Med' Adv' QUAD KNEE ("chimney") coil). Lower coil connect status refers to all PA coils / any coil using the lower port at the "dog house". Coil code should match that of what was detected (in this case the zeros were config'd as O and o (oh's - upper and lower case), and the number one following "ETApplies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.