Pular para o conteúdo principal
GE Healthcare · MRExcite22246662248108

DRF Over Range Errors Due to Gradient Power Supply

Symptom

Unable to scan with DRF Over Range Errors APS received data over range errors from DRF2 on all channels (0-7), error number 2248108. Shortly after, there may or may not be an error indicating the master gradient failed to go to ready when commanded.

Error message

2224666 Auto Prescan failed, Master Gradient is not ready

2248108 APS received data over range error from DRF, on chan= n. 2248108 TOiMR2 NSP APS:dataMoverTask 604 Exception Class: Idle Severity: Unknown Function: Acquisition/Pulse Generation: Acquisition PSD: Spin Echo Coil: US SHOULDER PA Scan: 379/4 APS received data over range error from DRF, on chan = 2 Need to reduce receiver gain.

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2334862-2 ACGD SGA PS ASM W/FRU PACKAGING See Long Description 2334862-2 Part Type: GP REPAIRABLES Part Weight: 2.18 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ACGD SGA PS ASM W/FRU PACKAGING FRU: Yes Status: Orderable This supply is for use with HFD, HFD-S, and ACGD Cabinets. ** WARNING - POSSIBLE PERSONAL INJURY. HEAVY OBJECT - THE POWER SUPPLY WITH CASE WEIGHS APPROXIMATELY 240 LBS (109 KG). Assistance to lift this product mechanical (Dollies) or Additional personnel] is required. 212002950 May 27, 2009
Area
MR