Pular para o conteúdo principal
GE Healthcare · X-ray R&FAdvantx

DRS Gateway loss of image data when patient entered from FUJI PACS worklist

Symptom

System configuration is Advantx Legacy with DRS / DRS Gateway using FUJI worklist broker. If a patient is entered on the DRS from the FUJI worklist broker, the DRS appears to acquire images normally. The Patient directory shows the expected number of images acquired in the run. When the images are displayed, they show up as black circles (No Image data). It is impossible to recover the images. If the New Patient data is entered manually without using the worklist broker, the DRS works normally.

Applies to

Models
Advantx
Product
DRS Gateway DRS
Area
X-ray R&F

Full cause & step-by-step resolution