Customer or Research Surface Coils Fail Coil Config After Upgrading from HD to HDx
Symptom
Customer or Research Surface Coils Fail Coil Config After Upgrading from HD to HDx After upgrading the 3T HD MR system to HDx platform a custom Cardiac Coil installed on the 12x system did not upgrade properly. Coil is supplied by GE ASL and is named "RM:6 Carotid"
Error message
2223625 Send packet opcode:30011 seqnum:11 rev:2 type:1 length:0 req:SCAN_01_NOR resp:COILCK status:0 failed. 2223625 lx-mr scn mrmailmgr.c 791 Exception Class: Unknown Function: Software/Firmware: Host Communications Send packet opcode:30011 seqnum:11 rev:2 type:1 length:0 req:SCAN_01_NOR resp:COILCK status:0 failed. 2223621 Internal Software Error. Can't callback on undefined NMRID:COILCK. 2223621 lx-mr scn mrmail_connect_mgr.c 239 Exception Class: Unknown Function: Software/Firmware: Host Communications Internal Software Error. Can't callback on undefined NMRID:COILCK. 2258113 Host coil validation error: Default Config Bias(defaultBiasCfg) field not found in CoilConfig.cfg. Connector: Multicoil Lower Port Coil: 8NVHEAD_A. Pls call your service representative 2258113 TRMR01 coilCK coilCK_scp.c 910 Host coil validation error: Default Config Bias(defaultBiasCfg) field not found in CoilConfig.cfg. Connector: Multicoil Lower Port Coil: 8NVHEAD_A. Pls call your service representative
Applies to
- Models
- Excite
- Product
- 3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
- Software
- 14.0 For both Excite HDx and 3T Excite HDx systems.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.