Bright White Line Artifact Collapsed Image Due to STIF or SRF/TRF I/O Board

Symptom

Collapsed image in H to F Direction on an Axial See Attachment. This image shows an apparent Z-axis collapse. The image was an axial Head.

Error message

2248108 APS received data over range error from DRF, on chan= n. 2248108 TOiMR2 NSP APS:dataMoverTask 604 Exception Class: Idle Severity: Unknown Function: Acquisition/Pulse Generation: Acquisition PSD: Spin Echo Coil: US SHOULDER PA Scan: 379/4 APS received data over range error from DRF, on chan = 2 Need to reduce receiver gain.

2253850 Gradient Link Hammer test failed. The GP received an invalid %c axis data value via the gradient data link. Check the gradient data and clock fiber optic cable connections 2253850 Gradient Link Hammer test failed. The GP received an invalid %c axis data value via the gradient data link. Check the gradient data and clock fiber optic cable connections

2244574 The GP reported a Gradient Amplifier % Axis Internal Wiring Fault or Control Undervoltage

2250082 Z Axis Gradient Data Overrange Occurred

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2280952-3 STIF Board See Long Description SRF TRF I/O or STIF 2280952 crosses up in FEMC to 2280952-3. Part Description is "STIF Assembly".
Area
MR