GE Healthcare · MRExcite
Intermittent zipper artifact and getting a diamond pattern in the EPT white pixel data
Symptom
MRIQ Intermittent zipper artifact and getting a diamond pattern in the EPT white pixel data Customer was getting intermittent zipper artifacts on various images, very intermittent. EPT White pixel noise test failed intermittently in head and body with the usual bright spots, but it also had a diamond pattern in the background. Image 258 shows what EPT was seeing, Image 264 was captured when using a sniffer, pattern is easier to see in image 264.
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- SecondOpinion 1.5T Excite HD 16 Channel Switch 5109645 See comments in resolution concept for second opinion requirements before ordering this part
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.