GE Healthcare · MRExcite2248277
Prescan Keeps Running Will Not Stop Due to APS Board
Symptom
Prescan Keeps Running Will Not Stop
Error message
2248277 APS did not receive dataRdyPkt from DRF Thu Aug 31 09:14:31 2006 Error: 2248277 Host: DMR2-OC0 Process: NSP File: APS:dataMoverTask Line: 655 Software error. APS did not receive dataRdyPkt from DRF Please perform a TPS Reset. frameCountPreScan = 15
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2294300-4 APS BOARD WITH IT DAUGHTER BD
- Area
- MR
Full cause & step-by-step resolution