GE Healthcare · MRExciteSigna
Calibrating the Excite 3T MKS RF Amplifier. 3T Excite G3.
Symptom
RF Calibration Notes for the MKS Amplifier All Signa EXCITE 3.0T Systems During the MKS RF Amp calibration, attenuation/gain is changed by using a keypad at the front of the amplifier. If the MODE button is pressed while RF power is active, the Body Hybrid or the Head T/R Switch can be damaged. SMILe 5/24/2004
Applies to
- Models
- Excite, Signa
- Product
- 3T Excite G3 Software ONLY
- Software
- G3 For 3T Excite only
- Area
- MR