GE Healthcare · MRExcite22487632254261
3T RF Amplifier Error - Prescan fails with no signal. 3T RF Amplifier front panel filter clogged with dust.
Symptom
3T RF Amplifier Error - Prescan Fails with No Signal No signal on any coil. 3T RF Amplifier errors 2248763 and 2254261.
Error message
2248763 RF Amplifier Error RF amplifier failed to power off in the allotted amount of time when commanded to do so 2248763 RF Amplifier Error. RF amplifier failed to power off, in the allotted amount of time, when commanded to do so. Please power cycle the RF Subsystem and perform a TPS Reset. 2254261 3T RF Amplifier PA Air Too Hot (fault 47). See Long Description Mon Oct 9 13:20:16 2006 Error: 2254261 Host: MR3T Process: NSP File: SCP:../rf_nb_contro Line: 1491 3T RF Amplifier PA Air Too Hot (fault 47). Possible Causes: obstructed air flow, PSD exceeding safe operating range. Cool 30 minutes, clear air intake and exhaust port, adjust PSD parameters. If the fault persists, replace the RF amplifier. Extended Error : The RF Amp PA Tube experienced high air temperature. Allow the RF Amp to cool for 20 minutes and restart scanning. Make sure room air temperature is <40 C, address with site if necessary. Check air filters and replace if necessary. If fault persists, replace the RF deck.
Applies to
- Models
- Excite
- Product
- 3T Excite HD 12.0 software only
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.