There is an error message logged indicating the "Master Gradient", or "Slave Gradient" but this is an HD system.
Symptom
There is an error message logged indicating the "Master Gradient" or "Slave Gradient", but this is an HD or HDxT system. There is no Master or Slave Gradient on an HD or HDxT system. See Long Description Why does this error message reference the "Master Gradient", or "Slave Gradient"? HD or HDxT systems have either HFD gradients or HFD-S gradients, with only one HFA per axis, e.g. there is no such thing as a "Master Gradient" or 'Slave Gradient" on HD and HDxT systems.
Error message
2248645 Master Gradient is not ready. Error: 2248645 Host: OMR2_OC0 Process: NSP File: APS:aps_pass2 Line: 238 Master Gradient is not ready. 2252239 Startup Timeout was detected on the Master Gradient. Mon May 23 22:49:44 2005 Error: 2252239 Host: signa2 Process: NSP File: SCP:CanAutoConfig Line: 1840 Startup Timeout was detected on the Master Gradient. 2252231 Network Management Recovery detected on the CAN network. Recovered Node is Master Gradient
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.