Pular para o conteúdo principal
GE Healthcare · MRExcite224864522522392252231

There is an error message logged indicating the "Master Gradient", or "Slave Gradient" but this is an HD system.

Symptom

There is an error message logged indicating the "Master Gradient" or "Slave Gradient", but this is an HD or HDxT system. There is no Master or Slave Gradient on an HD or HDxT system. See Long Description Why does this error message reference the "Master Gradient", or "Slave Gradient"? HD or HDxT systems have either HFD gradients or HFD-S gradients, with only one HFA per axis, e.g. there is no such thing as a "Master Gradient" or 'Slave Gradient" on HD and HDxT systems.

Error message

2248645 Master Gradient is not ready. Error: 2248645 Host: OMR2_OC0 Process: NSP File: APS:aps_pass2 Line: 238 Master Gradient is not ready.

2252239 Startup Timeout was detected on the Master Gradient. Mon May 23 22:49:44 2005 Error: 2252239 Host: signa2 Process: NSP File: SCP:CanAutoConfig Line: 1840 Startup Timeout was detected on the Master Gradient.

2252231 Network Management Recovery detected on the CAN network. Recovered Node is Master Gradient

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues