GE Healthcare · MRExcite225231851384
unable to scan dd/tr driver faults.
Symptom
Cannot Scan With intermittent TR and Dynamic Disable Faults. The customer when scanning is getting intermittent TR and Dynamic Disable faults. It happens with all coils and cannot be pinned down to a specific PSD.
Error message
2252318 Exception Class The Driver Module Has Detected a Power Supply Fault Source of Fault is High Voltage Power Supply. Fri Apr 7 08:18:07 2006 Error: 2252318 Host: b5 Process: NSP File: SCP:DCB Line: 1355 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: Vasc TOF Coil: HEAD Scan: 51384/3 The Driver Module Has Detected a Power Supply Fault. Source of Fault is High Voltage Power Supply. Actual Voltage = 10.0 Limit Voltage = 950.0->
Applies to
- Models
- Excite
- Product
- 3T Excite HD 12.0 software only
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2208999-7 3.0T RF Body Coil and Case
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.