Cannot scan in Head Transmit Mode due to Head TR Driver Faults.

Symptom

Cannot scan in Head Transmit Mode with Head TR Driver Faults. Body Transmit works OK.

Error message

2252293 The Driver Module Has Detected a Fault. Wed Apr 21 20:00:24 2004 Error: 2252293 Host: lx-mr Process: NSP File: SCP:DCB Line: 1986 The Driver Module Has Detected a Fault.^A Details of the Fault can be found in other error messages.->

2224667 Manual Prescan failed, Driver Module error occurred. 2224667 CMR1_OC0 psc PscMainControl.cc 1320 Manual Prescan failed, Driver Module error occurred.

2252305 TR Driver Fault Detected Open Circuit. Source of Fault is Head TR 2252305 lx-mr NSP SCP:DCB 1240 TR Driver Fault Detected - Open Circuit. Source of Fault is Head TR. Value: 0 mA Limit: 900 mA->

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR

Full cause & step-by-step resolution