MKS 3T Amplifier will not go to ready after several power outages at the site.

Symptom

MKS 3T Amplifier will not go to ready after several power outages at the site. What to Check? When powering the amplifier on you get a fault indication and then the amp goes to an off state.

Error message

2254253 3T RF Amplifier PA Filament under Voltage (fault 38). The PA Filament supply output was below specification(15V - 5%). 2254253 NSP SCP:../rf_nb_contro 1460 memp 8HRBRAIN protocol:DQA_8ch_Brain/1 3T RF Amplifier PA Filament under Voltage (fault 38). The PA Filament supply output was below specification(15V - 5%). Possible Causes: Defective supply or short on supply output. If the fault persists, replace the RF amplifier

Applies to

Models
Excite
Product
3T Excite G3 Software ONLY
Software
G3 For 3T Excite only
Component
SecondOpinion MKS 3T 35kW RF Amp 5104896 See comments in resolution concept for second opinion requirements before ordering this part.