GE Healthcare · MRExcite60642

Issue with Low Level Drift and 12x software

Symptom

Low Level Drift FMI 60642 Software will Not Work with Software Release 12x M4 For sites with FMI 60642 Low Level Drift, the software will not work with 12.0 M4 release. Not many sites have this FMI installed. E-mail from MR Engineering on September 1, 2006: "An issue was discovered recently where the Low Level Drift software will not work with release 12x_M4. The issue was resolved with 12x_M5". Proprietary, For GE Healthcare personnel Only For sites with FMI 60642 Low Level Drift, the software will not work with 12.0 M4 release. Not many sites have this FMI installed. Email from MR Engineering on September 1, 2006: An issue was discovered recently where the Low Level Drift software will not work with release 12x_M4. The issue was resolved with 12x_M5. Bottom line, If you have LLD, and 12x software, you must perform the M5 software FMI.

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR