GE Healthcare · MRExcite2218641

Unable to scan due to "SGD Option key status" error

Symptom

Unable to Scan with SGD Option key Status See Long Description This system is an Excite II Echospeed Gold Seal running 11.0 M4 software.. The system has ACGD gradient amplifiers. When starting installation calibration, the FE discovered the system wouldn't scan due to the error message shown below. He checked that the system is configured as an Echospeed and the GradientConfig.cfg file is correct. Also, the system has the "ACGD plus" option. What's wrong? Error: 2218641 Host: lx-mr Process: scn File: patCallbacks.c Line: 1070 SGD Option key status does not match the /w/config/GradientConfig.cfg SRMode. SRmode=120, GradAmpType=8915. Please correct configuration, and perform a TPS Reset.

Error message

2218641 SGD Option key status does not match the /w/config/GradientConfig.cfg SRMode. SRmode=120, GradAmpType=8915. Error: 2218641 Host: lx-mr Process: scn File: patCallbacks.c Line: 1070 SGD Option key status does not match the /w/config/GradientConfig.cfg SRMode. SRmode=120, GradAmpType=8915. Please correct configuration, and perform a TPS Reset.

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Area
MR

Full cause & step-by-step resolution