GE Healthcare · MR1765029850
Preposterous Y-Gradcal cfyfull Change Recommended during PM SPT
Symptom
Large Value Recommended in SPT Gradcal for cfyfull (Y-axis), about double expected value, and the measured length is 59.1 mm, we expect 100 mm See Long Description for full details During the y-axis gradcal portion of SPT it recommended a change for cfyfull from 17650 to 29850. The measured length is 59.1 mm. There is no real bubble in the phantom. Diagnostics passed. The recommended change is beyond the ability of the gradient.
Applies to
- Product
- Open Speed with EXCITE For HFO4 and HFO5 only
- Software
- OPENMR4-HFO4 Software version for HFO4 system, which is OpenSpeed with Excite Technology
- Component
- 2334383 P9438EB CONTROL BOARD ASSEMBLY SGA LITE Control Board
- Area
- MR
Full cause & step-by-step resolution