GE Healthcare · MR1765029850

Preposterous Y-Gradcal cfyfull Change Recommended during PM SPT

Symptom

Large Value Recommended in SPT Gradcal for cfyfull (Y-axis), about double expected value, and the measured length is 59.1 mm, we expect 100 mm See Long Description for full details During the y-axis gradcal portion of SPT it recommended a change for cfyfull from 17650 to 29850. The measured length is 59.1 mm. There is no real bubble in the phantom. Diagnostics passed. The recommended change is beyond the ability of the gradient.

Applies to

Product
Open Speed with EXCITE For HFO4 and HFO5 only
Software
OPENMR4-HFO4 Software version for HFO4 system, which is OpenSpeed with Excite Technology
Component
2334383 P9438EB CONTROL BOARD ASSEMBLY SGA LITE Control Board
Area
MR

Full cause & step-by-step resolution