Pular para o conteúdo principal
GE Healthcare · MR22445752245065496972244578224505049682

Unable to scan due to X-Axis Current Distortion & Overcurrent Faults due to X-axis SGA-LITE module

Symptom

Unable to Scan with X-Axis Current Distortion & Overcurrent Faults Unable to scan due to X-Axis Current Distortion & Overcurrent Faults. Cycling power and rebooting doesn't help. See Long description for a copy of the logged errors. Tue Aug 22 12:40:05 2006 2244578 WRMR1 NSP SCP:GP 1749 The GP is reporting a Gradient Amplifier X Axis Current Distortion Fault.-> Tue Aug 22 12:40:07 2006 2244575 WRMR1 NSP SCP:GP 1642 The GP is reporting a Gradient Amplifier X Axis OverCurrent Fault.-> Tue Aug 22 12:40:08 2006 2244603 WRMR1 NSP SCP:GP 1657 The GP reported multiple Gradient Amplifier X Axis OverCurrent Fault.-> Tue Aug 22 12:41:22 2006 2224666 WRMR1 psc PscMainControl.cc 1351 Auto Prescan failed, Gradient amplifiers are not ready

Error message

2244575 The GP is reporting a Gradient Amplifier X Axis OverCurrent Fault Error: 2244575 Host: WMR2 Process: NSP File: SCP:Grad Line: 2683 The GP is reporting a Gradient Amplifier X Axis OverCurrent Fault.->

2245065 GRADIENT DRIVER TESTS Static Test executed on the X axis since a fault was detected 2245065 WRMR1 diagCoord grad_lib.c 877 Diagnostic: GDT Stat Dyn Iteration: 1 of 1 Action: 1 Task ID: 0x3d5c638 Time: 49697 Object: SCP Test Task GRADIENT DRIVER TESTS - Static Tests will not be executed on the X axis since a fault was detected. Check previous error messages. To test this axis, clear the fault and run test again.

2244578 The GP is reporting a Gradient Amplifier % Axis Current Distortion Fault

2245050 GRADIENT DRIVER TESTS - %c Axis Fault Occurred The GP reported a %c Axis SGA UnderVoltage Fault 2245050 "GRADIENT DRIVER TESTS - %c Axis Fault Occurred.\n\ The GP reported a %c Axis SGA UnderVoltage Fault.\n\ If no other faults reported, check cables or replace\n\ the SGA. Else, see other fault error message.\n\ 2245050 WRMR1 diagCoord grad_lib.c 877 Diagnostic: GDT Stat Dyn Iteration: 1 of 1 Action: 1 Task ID: 0x3d5c638 Time: 49682 Object: SCP Test Task GRADIENT DRIVER TE

Applies to

Product
Open Speed with EXCITE For HFO4 and HFO5 only
Software
OPENMR5-HFO5 Open Speed with Excite Product
Component
2363414 SGA-LITE WITH FRU BOX P9438AB SGA-LITE WITH FRU BOX 2363414 P9438AB SGA-LITE WITH FRU BOX 2363414
Area
MR

Full cause & step-by-step resolution