GE Healthcare · MRExcite22538252253829
SSM replacement procedures not completed.
Symptom
Cannot Scan With intermittent TR and Dynamic Disable Faults. The customer when scanning is getting intermittent TR and Dynamic Disable faults. It happens with all coils and cannot be pinned down to a specific PSD.
Error message
2253825 CPD Board Failure in SSM chassis Command Register error 2253825 Diagnostic: TR DynamicDis Iteration: 1 of 1 Action: 1 CPD Board Failure in SSM chassis Command Register error. Check the SSM CPD Board. Register address: 20H Data read: 8H Expected result: 0H-> 2253829 TR/DD Failure in SSM chassis The service mode switch(s) on the front panel of the SSM are in the wrong position Thu Jun 8 15:45:52 2006 Error: 2253829 Host: lx-mr Process: diagCoord File: TrDiags.c Line: 387 Diagnostic: TR DynamicDis Iteration: 1 of 1 Action: 1 TR/DD Failure in SSM chassis The service mode switch(s) on the front panel of the SSM are in the wrong position. Configure the switches for normal operation.->
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.