GE Healthcare · MRExcite
Cradle Is Obstructed Due to Split Bridge Assembly
Symptom
Cradle Will Not Move In or Out Table raised normally. LPCA will not drive in or out.
Error message
The Cradle is Obstructed This is a popup error message that the operator sees when the cradle is commanded to move and the system does not sense the movement. (The system counts encoder pulses per unit of time and will fail if none or too few pulses are seen). The error message is not logged in the message log.
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution