44625 Image Disk Space Erroneous

Symptom

Image Disk Space Erroneous Customer reports that image disk space shows 80% +. If you perform a df, it reports the /export/home1 partition usage less than 15%. They should both be very close to the same.

Error message

Unable to allocate image space Following a database reset, the FE attempted to scan. Autoprescan failed with the error, "Unable to allocate image space"

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2368697-4 LINUX DDR WORKST'N 36GB 15K RPM 3.5 2368697-4 HP 36 GByte Hard Disk Drive 2368697-4 HP 36 GByte Hard Disk Drive