GE Healthcare · MRExcite22523022252293
Invivo HD knee coil is not compatible with HDx LPCA
Symptom
Some Early Units of Invivo HD Knee Coil Not Compatible with HDx LPCA See Long Description It has been identified that early units of the HD Knee Coil by Invivo (MRID) do not meet the design requirements for HDx and will not be compatible with forward production HDx(14.x) systems. The coil does not have DC blocks on RF, causing excessive current draw on RF.
Error message
2252302 Multicoil Bias Fault - Transmit Fault. 2252302 Multi-coil bias fault - transmit fault. MCD1 channel 3 of 16, counting from 1 Value: -72mA. Limit 0mA Mask of Failed channels: 0x9c MCD1 1-8 : A or C MCD1 9-16 : B MCD2 1-8 : B or Legacy MC MCD2 9-16 : C 2252293 The Driver Module Has Detected a Fault. Wed Apr 21 20:00:24 2004 Error: 2252293 Host: lx-mr Process: NSP File: SCP:DCB Line: 1986 The Driver Module Has Detected a Fault.^A Details of the Fault can be found in other error messages.->
Applies to
- Models
- Excite
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 For both Excite HDx and 3T Excite HDx systems.
- Component
- SecondOpinion 1.5T 8 Channel TR Knee Array 5114356-2 See comments in resolution concept for second opinion requirements before ordering this part See comments in resolution concept for second opinion requirements before ordering this part
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.