GE Healthcare · MRSignaOvationExcite2236610
44397 Unable to Download or Reset The TPS MGD or ISE
Symptom
Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis
Error message
2236610 Memory test failed on CAP SRAM.This is a Fatal error. CERD boot is aborted. Error: 2236610 Host: lx-mr Process: NSP File: TPS:init1.c Line:214 CERD Diags: Memory test failed on CAP SRAM.This is a Fatal error. CERD boot is aborted. If the error can be reproduced, the CERD board needs to be replaced Address : 0000226cH Expected : 00008000H Actual : 00008001H
Applies to
- Models
- Signa, Ovation, Excite
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.