Will Not Scan With RF Errors. 2254278 3T RF Amplifier Enable Line fault 93.

Symptom

Will Not Scan With RF Errors Will not scan with RF errors

Error message

2224666 Auto Prescan Failed, RF Amplifier is Not Ready NOTE: When using this error to create new knowledge, please include any associated error codes. 2224666 is a generic error code which requires further information. Mon May 17 08:19:08 2004 Error: 2224666 Host: lx_mr Process: psc File: PscMainControl.cc Line: 1296 Auto Prescan failed, RF amplifier is not ready.

2254278 3T RF Amplifier Enable Line fault 93. The RF amplifier indicates that it was unblanked continuously for an excessive amount of time. Error: 2254278 Host: CENTMR02 Process: NSP File: SCP:../rf_nb_contro Line: 1491 3T RF Amplifier Enable Line (fault 93). The RF amplifier indicates that it was unblanked continuously for an excessive amount of time. Check PSD parameters. If the fault persists, replace the RF amplifier.

Applies to

Models
Excite
Product
3T Excite G3 Software ONLY
Software
G3 For 3T Excite only
Component
5112441 4KW MNS 3T RF AMP
Area
MR