Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova120011413001141200111130111

43466 Innova System boot failure

Symptom

Innova System boot failure Any defect that prevents from getting the first acquisition after the system reset. Customer needs to boot again to recover. When Innova system has a actor issue.

Error message

INx100: 1200114 ERR_DETECTOR_BOOT_WARNING Debug: RX read timeout (xxx ms), RX IT never received 1200114 ERR_DETECTOR_BOOT_WARNING DO XRImDet FPGA_Driver::update Detector_warning 1 1 [FPGA_Driver::update] Detector com error: CDetectorDipb::waitRxCommandDm RX read timeout (350 ms), RX IT never received, 0 RX Command read (last one is: 0 0) fpga_driver.cpp 541

error Code = 1200114, or (1300114 BiPlane) error Text = ERR_DETECTOR_BOOT_WARNING followed by error Code = 1200111,(130111 for BIPLANE) error Text = ERR_DETECTOR_BOOT_FAILURE System impact: Boot failure. Displayed message : "Xray aborted - Restart Xray; if problem persists, call Service". error Code = 1200114 or (1300114 BiPlane), error Text = ERR_DETECTOR_BOOT_WARNING , debug Text += Detector com error: CDetectorDipb::waitRxCommandDm RX read timeout (, recovery Class = DO, major Function /= XRImDet followed by error Code = 1200111, error Text = ERR_DETECTOR_BOOT_FAILURE, debug Text /= , recovery Class = IH, major Function /= XRImDet System impact: Boot failure. Displayed message : "Xray aborted - Restart Xray; if problem persists, call Service". error Code = 1200114 or (1300114 BiPlane), error Text = ERR_DETECTOR_BOOT_WARNING 

Applies to

Models
Innova
Product
Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100 Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System
Component
Detector and Detector Power Supply Detector PS 2302361-2 for 21 cm detector, 2302361-3 (alt # 2300475-2) for 31/41 detector Detector PS 2302361-2 for 21 cm detector, 2302361-3 (alt # 2300475-2) for 31/41 detector
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.