Pular para o conteúdo principal
GE Healthcare · MRExciteSigna224810210151224800622481212248131

Customer cannot scan due to RF errors.

Symptom

Will Not Scan With RF Errors Will not scan with RF errors

Error message

2248102 Received Invalid cm Packet 10151, scan state = 3 Tue May 4 20:27:03 2004 Error: 2248102 Host: lx-mr Process: NSP File: APS:processPkt Line: 896 Received invalid cm packet 10151, scan state = 3 Software error, please perform a TPS Reset.

2248006 DRF CM DMA Resource Not Available Tue May 4 20:27:03 2004 Error: 2248006 Host: lx-mr Process: NSP File: APS:DRF Line: 229 DRF CM DMA resource not available. This happens when DRF receives a filter select before it finishes acquiring data.

2248121 APS Send DRF X cmd Message Error (no response) Tue May 4 20:27:03 2004 Error: 2248121 Host: lx-mr Process: NSP File: APS:DrfDriver::sendCmdMsg Line: 647 APS Send DRF 1 cmd message error (no response) DRF is not running, please perform a TPS Reset.

2248131 APS DC Send DRF info error -1 Tue May 4 20:27:03 2004 Error: 2248131 Host: lx-mr Process: NSP File: APS:DacqControl::processPkt Line: 536 APS DC send DRF info error -1 Software error, please perform a TPS Reset.

Applies to

Models
Excite, Signa
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues