3T Body SPT Fails after Phantom Positioning Check Due to Phantom Holder
Symptom
3T HDx Body SPT fails after phantom positioning check is run 3T HDx upgrade completed. When Body SPT is run, SPT fails and the message displayed is Wrong Phantom size detected. Confirm 27cm Phantom is being used. Head SPT runs fine.
Error message
Wrong Body TLT phantom selected Use 27cm body phantom and loader No error is placed in the error log. It is displayed in SPT window.
Applies to
- Models
- Excite
- Product
- 3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
- Software
- 14.0 For both Excite HDx and 3T Excite HDx systems.
- Component
- 3T Body TLT Phantom Loader Shell 2360037
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.