GE Healthcare · MRExcite

3T Body SPT Fails after Phantom Positioning Check Due to Phantom Holder

Symptom

3T HDx Body SPT fails after phantom positioning check is run 3T HDx upgrade completed. When Body SPT is run, SPT fails and the message displayed is Wrong Phantom size detected. Confirm 27cm Phantom is being used. Head SPT runs fine.

Error message

Wrong Body TLT phantom selected Use 27cm body phantom and loader No error is placed in the error log. It is displayed in SPT window.

Applies to

Models
Excite
Product
3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
Software
14.0 For both Excite HDx and 3T Excite HDx systems.
Component
3T Body TLT Phantom Loader Shell 2360037
Area
MR

Full cause & step-by-step resolution