GE Healthcare · MRExcite222435822520631095367932
APS motherboard and IT link failures
Symptom
APS motherboard and IT link failures causing Missing Images This error shown in the log: unable to open file /export/home1/sdc_image_pool/reserved/IMSseR7C8Ye/5754.4.1.5.sdcopen. Error = 2. When this error occurred it was coincident with a missing image on the disk and in the image series just scanned. Recover database did not help, nor did reloading software. Ran MGD Host >> IT >> APS datapath diagnostics and had 3 failures in 100 reported. Error code was 2252063. Replaced the APS board with IT daughterboard. Diagnostics passed.
Error message
2224358 Unable to open image file 2224358 lx-mr TIR update_image_header.c 223 Unable to open image file /export/home1/sdc_image_pool/reserved/IMSseSAA a000V5/8407.2.1.36. Errno = 2 2252063 The test data transferred over the IT bus contains incorrect values. Thu Sep 16 16:52:12 2004 Error: 2252063 Host: lx-mr Process: diagCoord File: bit3_dp.c Line: 233 Diagnostic: Host->IT->APS Iteration: 1 of 1 Action: 1 Task ID: 0x0 Time: 1095367932 Object: Host Diag Executive The test data transfered over the IT bus contains incorrect values. This might be caused by bad IT bus boards
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.