GE Healthcare · MRExciteSigna224631910862442513075322482172847174
RRF Calibration Error in RRF Data Path Diagnostics
Symptom
RRF Calibration Error in RRF Data Path Diagnostics
Error message
2246319 DRF1 failed the DSP Self Test.-> see long description Error seen with MGD diags: 1086244251 0 1 Thu Jun 3 06:30:51 2004 2246319 2 qstmr1 diagCoord DrfDiags.c 954 Diagnostic: DRF 1 BLD Iteration: 1 of 1 Action: 1 Task ID: 0x3a8d2d8 Time: 30753 Object: APS Test Task DRF1 failed the DSP Self Test.-> 2248217 Receiver calibration error(s). Receiver DB out of spec. See Long Description 2248217 4b2a diagCoord RrfCalibrator::rcvCal 3779Diagnostic: RRF Calib Iteration: 32 of infinite Action: 1Task ID: 0x9d29e0 Time: 2847174 Object: APS Receiver calibration error(s). Receiver DB out of spec. R1 values that failed: Rx1: 12 10 Rx2: 12 11 10 9 Rx3: 12 11 10 9 Rx4: 12 11 10 Rx5: 12 11 10 9 Rx6: 12 11 10 9 Rx7: 12 11 10 9 Rx8: 12 10
Applies to
- Models
- Excite, Signa
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.