GE Healthcare · MRExciteDiscoveryOptima22572102248679
CWA ID 40485 Unable to Start a New Patient Entry.
Symptom
Unable to Start a New Patient Entry. This site had crashed numerous times as a result of a severe storm. After a successful reboot, and TPS RESET, the new patient entry was not responding and the system was reporting messages of Startup of Application failed, Reset TPS and MDS Link failure appears to be broken between SCP and the Gradient.
Error message
2257210 SCP did not respond to the start packet of WavRec. Mon Nov 22 10:57:41 2004 Error: 2257210 Host: MRI1-OC0 Process: WavRec File: WavRec.cc Line: 175 SCP did not respond to the start packet of WavRec. This may be due to a TPS Reset 2248679 RF Cabinet MDS peripheral response timeout Tue Jun 6 06:20:12 2006 Error: 2248679 Host: lx-mr Process: NSP File: SCP:MdsAutoConfig Line: 1830 RF Cabinet MDS peripheral response timeout. Unknown link failure internal to the RF Cabinet. Unknown link: 0x00 Please power cycle the RF Subsys
Applies to
- Models
- Excite, Discovery, Optima
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.