Pular para o conteúdo principal
GE Healthcare · MRExcite224530916278472245233112849511282941128394

Overnight Diagnostics Consistently Fail Gradient Driver Tests ECC Verification Test 2245233

Symptom

Overnight Diagnostics Consistently Fail See Error Concept for the diagnostics failures

Error message

2245309 Gradient Diagnostic test has been aborted because GP detected a bad clock input 2245309 WCLAGEMR2MR diagCoord grad_lib.c 924 Diagnostic: Grad Frame/Clk Iteration: 25 of infinite Action: 1 Task ID: 0x3fc4380 Time: 1627847 Object: SCP Test Task Function: Gradient: N/A Gradient Diagnostic test has been aborted because GP detected a bad clock input.-> Extended error log: "Check placement of Clock and X,Y,Z data fibers between the STIF and GP (look for swapped fibers between clock and x/y/z data fibers)" 2245309 WCLAGEMR2MR diagCoord grad_lib.c 924 Diagnostic: Grad Frame/Clk Iteration: 25 of infinite Action: 1 Task ID: 0x3fc4380 Time: 1627847 Object: SCP Test Task Function: Gradient: N/A Gradient Diagnostic test has been aborted because GP detected a bad clock input.-> Extended error log: "Check placement of Clock and X,Y,Z data fibers between the STIF and GP (look for swapped fibers between clock and x/y/z data fibers)"

2245233 GRADIENT DRIVER TESTS ECC Verification Test failed on Z Axis IERROR Gain out of range 2245233 WCLAGEMR2MR diagCoord grad_lib.c 924 Diagnostic: Grad ECC Verification Iteration: 7 of infinite Action: 1 Task ID: 0x3fc4380 Time: 1128495 Object: SCP Test Tas

Applies to

Models
Excite
Product
Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
Software
HDe 14.0 For Hino Excite HDe 14.0 systems only.
Component
5146875 GP3 for HDe
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues