High Incidence of Returned ACGD Power Supply With No Defect Found

Symptom

High Incidence of Returned ACGD Power Supply With No Defect Found Subject: MR SMILe: High incidence of ACGD Power Supply No Defect Found power supplies. Effectivity: All Signa systems with ACGD Power Supplies (GEHC p/n T3504MH, 2334862-2, 2317092-2, 2284857-2, 2272460-2 ). Problem Description: The Repair Development Center (RDC) has seen a high rate of No Defect Found field return power supplies during repair. SMILe - May 4, 2004

Error message

2244591 SGA Power Supply Over Voltage Fault See Long Description. Fri May 7 16:14:56 2004 Error: 2244591 Host: NSM6 Process: NSP File: SCP:GP Line: 1857 SGA Power Supply Over Voltage Fault. FOR FIELD ENGINEER: Look at SGA-PS front LEDs. If BUS OV LED lit, check PDU 420V output. If OV X,Y,Z LED lit, swap military connectors to isolate SGA vs. SGA-PS. Power cycle ACGD POWER breaker on PDU to recover.

Applies to

Models
Excite, Signa
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.