No Head or Multi Coil Signal and Body Signal is Good after upgrade to Excite HDx

Symptom

No Multi Coil Signal but Body Signal is Good after upgrade to Excite HDx HDX upgrade from 12X HD, body signal is good , but no signal in head or multicoil. MC Bias faults or no surface coil signal after HD (12.0) to HDx (14.0) Upgrade. See Long Description. Subject: MC Bias faults or no surface coil signal after HD (12.0) to HDx (14.0) Upgrade. Effectivity: All 1.5T and 3T EXCITE HDx (14.0) MR systems. Problem: System reports Multicoil Bias Faults after completing HD (12.0) to HDx (14.0) upgrade. Coils attached to the legacy Bendix connector will report multicoil bias open faults. Transmit/receive coils, like the Knee TR coil, connected to Hypertronics connector 'A' will scan without faults but no signal will be seen.

Applies to

Models
Excite
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 For both Excite HDx and 3T Excite HDx systems.
Area
MR

Full cause & step-by-step resolution