GE Healthcare · MRSignaOvationExcite2250081
38413 Gradient Data Overrange Errors and the Scan Completes with Good Images
Symptom
Gradient Data Overrange Errors and the Scan Completes with Good Images
Error message
2250081 Z Axis Gradient Data Overrange Occurred. The GP detected an error in the incoming High Speed Data from the STIF/SRF. Wed Jun 29 06:40:35 2005 Error: 2250081 Host: lx-mr2 Process: NSP File: SCP:GP Line: 1981 Z Axis Gradient Data Overrange Occurred.^A The GP detected an error in the incoming High Speed Data from the STIF/SRF. Swap fiber optics at the Systems Cabinet. Run fiber optic diagnostics.
Applies to
- Models
- Signa, Ovation, Excite
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution