GE Healthcare · MRExcite222384922521060400000067108864
Unable to ping AGP Board.
Symptom
Unable to Ping AGP Board System will not boot, unable to ping AGP board. The BFL light is orange and CPU is green on the AGP board.
Error message
2223849 AGP EPROM power-up boot failed See Long Description Error: 2223849 Host: rmns Process: NSP File: TardisReset.cc Line: 1781 AGP EPROM power-up boot failed. Communication Timed-out from AGP board. Verify that MCP820 boards are installed and seated properly for APS and AGP. Extended Error : MCP820 boards can be identified by the word MCP820 printed on the board extraction lever. 2252106 Failed to determine hardware state for AGP board within 20 seconds Error: 2252106 Host: rmns Process: NSP File: diags_utils.c Line: 666 Diagnostic software error. Failed to determine hardware state for AGP board within 20 seconds The board may be busy loading vxWorks or may be hung. RAM ADR: Addressability Running; FAILED While attempting to boot or download the TPS in a C-shell window type; mgd_term agp Messages will scroll during the boot. See Long Description for a sample of messages. Copyright Motorola Inc. 1988 - 2000, All Rights Reserved PPC1 Debugger/Diagnostics Release Version 4.9 - 07/10/00 RM01 COLD Start Local Memory Found =04000000 (&67108864) MPU Clock Speed =367Mhz BUS Clock Speed =67Mhz WARNING: Keyboard Not Connected Reset Vector Location : ROM Bank B Mezzanine Configuration:
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.