Intermittent severe corduroy on lumbar when cradle is extended more than 933 mm. on a Contour mobile.
Symptom
MRIQ Intermittent Corduroy Artifact on Lumbar Scans when the Cradle is Extended More Than 933 mm. on a Contour Mobile. Link to image quality troubleshooting guide. There are 3 cables which run through the cradle and are flexed when the cradle reaches out. The RX 0, RX 90 and Coil ID cables between the cradle end and preamp. When the cradle drives into the bore, the bend angle on the cables will decrease to around 45 degrees @ 933mm. This constant flexing will breakdown a cable. http://3.87.118.28/optec3/common/8xdocs/epst/mod11/m11_010_000.html
Applies to
- Product
- Contour 1 & 2 All Software Revs.
- Software
- 7.6x Profile1 through 3 and Contour1 through 3
- Component
- 2114578 Coil ID Cable 2114577 RX-90 Cable 2114576 RX-0 Cable
- Area
- MR
Full cause & step-by-step resolution