GE Healthcare · MRExcite222458322527392252716225008122527262252738
2224583 Auto and Manual Prescan Fails and 2250081 Gradient Data Overrange with Sequence playout errors
Symptom
Cannot scan an Abdomen sequence because of various TPS, Gradient and RF Amp errors.
Error message
2224583 Auto Prescan Failed, Possible Data Acquisition Failure. File: APS:aps_cf Line: 15 Auto Prescan failed, possible data acquisition failure. 2252739 Sequencer DSP RF_SEQ was active when synchronizer EOS occurred. 2252716 Sequence playout error The SRF synchronizer FIFO became empty 2250081 Z Axis Gradient Data Overrange Occurred. The GP detected an error in the incoming High Speed Data from the STIF/SRF. Wed Jun 29 06:40:35 2005 Error: 2250081 Host: lx-mr2 Process: NSP File: SCP:GP Line: 1981 Z Axis Gradient Data Overrange Occurred.^A The GP detected an error in the incoming High Speed Data from the STIF/SRF. Swap fiber optics at the Systems Cabinet. Run fiber optic diagnostics. 2252726 SRF error. One or both DSP sequencers on the SRF board has indicated an error 2252726 SRF error. One or both DSP sequencers on the SRF board has indicated an error. 2252738 Sequencer DSP SSP_GRD_SEQ did not receive data from AGP prior to SSI causing a synchronizer EOS error. Error: 2252738 Host: mrs1_oc0 Process: NSP File: AGP:seq_drv.c Line: 1686 Sequencer DSP SSP_GRD_SEQ did not receive data from AGP prior to SSI causing a synchronizer EOS error. The AGP did not have enough time to downl
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.