GE Healthcare · MRSignaOvationExcite2237285223729722372732227556
Unable to Download or Reset The TPS MGD or ISE with 2227556 and CERD Error 2237273.
Symptom
Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis
Error message
2237285 CERD BOOT FAILURE Unable to load software to the CERD board. 2237285 Host: TSMR Process: NSP File: TPS:cx_init.c Line: 117 CERD BOOT FAILURE Unable to load software to the CERD board. 2237297 CERD BOOT ERROR The CERD did not respond to a boot command in the allotted time. Host: lx-bay14b Proc: NSP Error: 2237297 File: TPS:cx_dpr_utils.c Line: 113 CERD BOOT ERROR The CERD did not respond to a boot command in the allotted time. CERD command : 32 response time : 100 2237273 CERD BOOT ERROR CERD boot software on TPS timed out while waiting for a response from the CAP. Host: lx-bay14b Proc: NSP Error: 2237273 File: TPS:cx_dpr_utils.c Line: 196 CERD BOOT ERROR CERD boot software on TPS timed out while waiting for a response from the CAP. IT MAY BE NECESSARY TO CYCLE POWER ON THE ISE CHASSIS TO CLEAR THIS CONDITION. timeout duration = 100 2227556 *** Reset/Download TPS was not successful. *** 2227556 sdm sdm_main.c 320 *** Reset/Download TPS was not successful. ***
Applies to
- Models
- Signa, Ovation, Excite
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.