GE Healthcare · MRExciteDiscoveryOptima22238492252106224715611409
Computer Will Not Boot With The Message On The Screen Says Do Not Interact With System
Symptom
Computer will not boot up to applications level, with the message on the screen that says Please do not interact with the system until this message disappears. Computer will not boot up to applications level, with the message on the screen that says -- > Please do not interact with the system until this message disappears. If the system does not come up in 15 minutes, please call your service representative.
Error message
2223849 AGP, SCP, APS EPROM power-up boot failed 2252106 Failed to determine hardware state for APS board within 20 seconds. Error: 2252106 Host: BARLOW_MR Process: NSP File: diags_utils.c Line: 666 Diagnostic software error. Failed to determine hardware state for APS board within 20 seconds The board may be busy loading vxWorks or may be hung. Also check the serial cable for the board and the serial terminal server. 2247156 The IRF2 board was not detected in the MGD chassis Error: 2247156 Host: BOWIE-MRI Process: diagCoord File: Irf.cpp Line: 74 Task ID: 0x3f1fdf8 Time: 11409 Object: APS The IRF2 board was not detected in the MGD chassis. The system will not function without this board.->
Applies to
- Models
- Excite, Discovery, Optima
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.