Driver Module not calibrated.

Symptom

Unable to Scan Head or Body with TR Driver Faults

Error message

2252357 Driver Module error occurred. 2252357 1 lx-mr NSP APS:aps_fastta 135 Driver Module error occurred.

2252310 DD Driver Fault Detected - Open Circuit See Long Description for extended error message. Fri Oct 22 19:03:04 2004 Error: 2252310 Host: PRD_GE_MR03 Process: NSP File: SCP:DCB Line: 1429 DD Driver Fault Detected - Open Circuit. Source of Fault is BD2 (Dynamic Disable Boards at Front of Body Coil). Value = 3644 mV Limit: 2177 mV-> Extended Error: This usually indicates that a cable somewhere in the circuit path is not connected or one of the load diodes has failed open. Inspect the failing path. Use TR DD MC System Path diagnostic to troubleshoot. In HD/HDx Systems, BD1 is the Dynamic Disable Board at Front of Body Coil BD2 is the Dynamic Disable Board at Front of Body Coil BD3 is the Dynamic Disable Board at Back of Body Coil BD4 is the Dynamic Disable Board at Back of Body Coil In 20.0 and later systems, try running the TR DD System Path diagnostic to troubleshoot.

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR