GE Healthcare · MRSignaOvationExcite22493911126778983137472249392
Fails TPS Reset when booting up to Applications or when Resetting TPS
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2249391 semTake failed for semaphore ID 0x3FFFDF0 1126778983 2 1 Thu Sep 15 10:09:43 2005 2249391 4 MR3T diagCoord sysIntegrity.c 1010 Task ID: 0x3fffbb0 Time: 13747 Object: APS semTake failed for semaphore ID 0x3FFFDF0. Errno = 0x3D0004. EN 218 SR 218 2249392 semGive failed for semaphore ID 0x3FFFDF0 Thu Sep 15 10:09:43 2005 2249392 4 MR3T diagCoord sysIntegrity.c 1027 Task ID: 0x3fffbb0 Time: 13747 Object: APS semGive failed for semaphore ID 0x3FFFDF0. Errno = 0x160068.
Applies to
- Models
- Signa, Ovation, Excite
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.