Symptom
Scan fails immediately with IT-Host card driver initialization error
Error message
2252055 Diagnostic: IT Host->MGD Hammer. IT-Host card driver initialization error. 2252055 lx-mr diagCoord bit3_diags.c 469 Diagnostic: IT Host->MGD Hammer Iteration: 1 of 1 Action: 1 Task ID: 0x0 Time: 1121724332 Object: Host Diag Executive IT-Host card driver initialization error. This error usually means that the IT-Cable is disconnected or that the cable is damaged.->
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution