GE Healthcare · MRExcite222362122572102223625101682249107

SCP/SRI link has failed attempt to recover Due to J13 reset fiber and J11 Xmit fibers on SRI3 were swapped Excite HD

Symptom

SCP/SRI Link has Failed Attempt to Recover Error on Scan page

Error message

2223621 Internal Software Error. Can''t callback on undefined NMRID:SPU. Mon Nov 22 10:57:41 2004 Error: 2223621 Host: MRI1-OC0 Process: WavRec File: mrmail_connect_mgr.c Line: 238 Internal Software Error. Can''t callback on undefined NMRID:SPU.

2257210 SCP did not respond to the start packet of WavRec. Mon Nov 22 10:57:41 2004 Error: 2257210 Host: MRI1-OC0 Process: WavRec File: WavRec.cc Line: 175 SCP did not respond to the start packet of WavRec. This may be due to a TPS Reset

2223625 Send packet opcode:10168 seqnum:7 rev:2 type:1 length:0 req:WavRec resp:SPU status :0 failed. Mon Nov 22 10:57:41 2004 Error: 2223625 Host: MRI1-OC0 Process: WavRec File: mrmailmgr.c Line: 791 Send packet opcode:10168 seqnum:7 rev:2 type:1 length:0 req:WavRec resp:SPU status :0 failed.

2249107 SRI SCP has recieved a packet indicating the SRI has reset SRI heart beat is steady on 2249107 SRI SCP has recieved a packet indicating the SRI has reset SRI heart beat is steady on

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR