GE Healthcare · MRExcite22522312249412
intermittently, there is no knocking...it just counts down as if scanning. customer reboots and is able to resume scanni…
intermittently, there is no knocking...it just counts down as if scanning. customer reboots and is able to resume scanning as normal.
Symptom
Acquisition Timer counts down but no Images are acquired Intermittently there is no knocking, timer counts down as if scanning. No images reconstruct. Customer reboots and is able to resume scanning as normal.
Error message
2252231 Network Management Recovery detected on the CAN network. Recovered Node is Drivers Control Board. Error: 2252231 Host: lx-mr Process: NSP File: SCP:CanLinkHWO Line: 663 Network Management Recovery detected on the CAN network. Recovered Node is Drivers Control Board. 2249412 A PCI error occured. The PCI Parity Error PERR was asserted. 2249412 MRS1_OC0 diagCoord excLog.c 1357 Task ID: 0x3dcdef0 Time: 6163 Object: SCP A PCI error occured. The PCI Parity Error PERR was asserted. Qspan PCI Status Reg: 0x8390; Qspan Interrupt Status Reg: 0x30408C00 Qspan Interrupt Control Reg: 0x30600000. Additional information in /usr/g/service/log/vxPciLg-16d9
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.