Pular para o conteúdo principal
GE Healthcare · MRExcite22522312249412

intermittently, there is no knocking...it just counts down as if scanning. customer reboots and is able to resume scanni…

intermittently, there is no knocking...it just counts down as if scanning. customer reboots and is able to resume scanning as normal.

Symptom

Acquisition Timer counts down but no Images are acquired Intermittently there is no knocking, timer counts down as if scanning. No images reconstruct. Customer reboots and is able to resume scanning as normal.

Error message

2252231 Network Management Recovery detected on the CAN network. Recovered Node is Drivers Control Board. Error: 2252231 Host: lx-mr Process: NSP File: SCP:CanLinkHWO Line: 663 Network Management Recovery detected on the CAN network. Recovered Node is Drivers Control Board.

2249412 A PCI error occured. The PCI Parity Error PERR was asserted. 2249412 MRS1_OC0 diagCoord excLog.c 1357 Task ID: 0x3dcdef0 Time: 6163 Object: SCP A PCI error occured. The PCI Parity Error PERR was asserted. Qspan PCI Status Reg: 0x8390; Qspan Interrupt Status Reg: 0x30408C00 Qspan Interrupt Control Reg: 0x30600000. Additional information in /usr/g/service/log/vxPciLg-16d9

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues