GE Healthcare · MRExcite22522312249412
intermittently, there is no knocking...it just counts down as if scanning. customer reboots and is able to resume scanni…
intermittently, there is no knocking...it just counts down as if scanning. customer reboots and is able to resume scanning as normal.
Symptom
Acquisition Timer counts down but no Images are acquired Intermittently there is no knocking, timer counts down as if scanning. No images reconstruct. Customer reboots and is able to resume scanning as normal.
Error message
2252231 Network Management Recovery detected on the CAN network. Recovered Node is Drivers Control Board. Error: 2252231 Host: lx-mr Process: NSP File: SCP:CanLinkHWO Line: 663 Network Management Recovery detected on the CAN network. Recovered Node is Drivers Control Board. 2249412 A PCI error occured. The PCI Parity Error PERR was asserted. 2249412 MRS1_OC0 diagCoord excLog.c 1357 Task ID: 0x3dcdef0 Time: 6163 Object: SCP A PCI error occured. The PCI Parity Error PERR was asserted. Qspan PCI Status Reg: 0x8390; Qspan Interrupt Status Reg: 0x30408C00 Qspan Interrupt Control Reg: 0x30600000. Additional information in /usr/g/service/log/vxPciLg-16d9
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.