Pular para o conteúdo principal
GE Healthcare · CTHiSpeedBrightSpeedVCT23000006450003164000245043245190

Scanning aborted WITH TAXI data violation errors

Symptom

"Scan hardware Stopped Scan" with Taxi data violation errors. "Scan hardware Stopped Scan" with Taxi data violation errors

Error message

Error: 230000064 TAXI data violation detected by forward error correction See the long description for the full error code Sat Apr 17 20:26:43 2004 Error: 230000064 Host: darc Process: dataacq File: dip_thread.cxx Line: 411 Running diagnostic: Scan Data Path Functional Function: Data Acquisition : Scan Data Acquisition Scan Type: Axial Scan: 50003/5/1 Scan Seq Id: 164000 TAXI data violation detected by forward error correction. Interrupt bit: 4 Violation count: 1 Scan aborted.

Error: 245043 Scanning System Aborted Scanning. See LD for complete error Tue Feb 7 07:50:12 2006 Error: 245043 Host: ctscan1 Process: scanmgr File: scanmgr_events.cxx Line: 2115 Function: System Control : Scan Control Scan Type: Axial Scan: 0/0/0 Scan Seq Id: 1000 Scanning System Aborted Scanning

Error: 245190 System aborted Axial scanning. No views, either offset or data, were collected/received. Data views expected: 992 See the Long Description for the full error code. Thu Jan 13 06:58:23 2005 Error: 245190 Host: ct02 Process: scanmgr File: scan_complete.cxx Line: 303 Function: System Control : Scan Control Scan Type: Axial Scan: 0/0/0 Scan Seq Id: 1000 System aborted Axial scanning. No views, either off

Applies to

Models
HiSpeed, BrightSpeed, VCT
Product
HiSpeed QX/i - GOC2 103L.5_HSQXIM4 103L.4_HSQXIM4 102L.1_HSQXIM3 HiSpeed QX/i - GOC2 Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types GOC2 List of softwares the product can run 103L.5_HSQXIM4 103L.4_HSQXIM4 - obsolete 102L.1_HSQXIM3 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues