Pular para o conteúdo principal
GE Healthcare · MRExcite2248321

System stops scanning and tech had to do a TPS reset frequently.

Symptom

System Stops Scanning and Images do not Reconstruct. Frequent TPS Reset needed. Note: Ping to APS, AGP & Terminal Server are okay. System lock up only happens during scans. Sometimes scanning is completed but no images come up in the Autoview and nothing listed in the Image Browser.

Error message

2248321 - The SC server is not responding, the server might be hung. Error: 2248321 Host: MR01 Process: NSP File: SCP:sct_wdog_exp Line: 1003 Communication Error. The callback for SCSM state 0x2 was not received in time. The SC server is not responding, the server might be hung. Please perform a TPS reset to recover.

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2272740-3 8 PORT 10/100 INTEL HUB SERVICE KIT 2272740-3 8 PORT 10/100 INTEL HUB SERVICE KIT. 2379186 Term Server
Area
MR

Full cause & step-by-step resolution