System stops scanning and tech had to do a TPS reset frequently.
Symptom
System Stops Scanning and Images do not Reconstruct. Frequent TPS Reset needed. Note: Ping to APS, AGP & Terminal Server are okay. System lock up only happens during scans. Sometimes scanning is completed but no images come up in the Autoview and nothing listed in the Image Browser.
Error message
2248321 - The SC server is not responding, the server might be hung. Error: 2248321 Host: MR01 Process: NSP File: SCP:sct_wdog_exp Line: 1003 Communication Error. The callback for SCSM state 0x2 was not received in time. The SC server is not responding, the server might be hung. Please perform a TPS reset to recover.
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2272740-3 8 PORT 10/100 INTEL HUB SERVICE KIT 2272740-3 8 PORT 10/100 INTEL HUB SERVICE KIT. 2379186 Term Server
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.