Under certain conditions swapping a faulty SGA or HFA Control Board into a functional gradient amplifier can damage it

Symptom

Gradient Amplifier could be damaged by a defective SGA or HFA Control Board SMILe 1/19/06 Effectivity: All Signa 1.5T and 3T MR Systems using ACGD, HFD or HFA gradients. Problem: It is possible, during troubleshooting, to swap a faulty SGA or HFA Control Board into a functioning amplifier and damage the amp. SMILe 1/19/06 Effectivity : All Signa 1.5T and 3T MR Systems using ACGD, HFD or HFA gradients. Problem: It is possible, during troubleshooting, to swap a faulty SGA or HFA Control Board into a functioning amplifier and damage the amp.

Error message

2244576 The GP is reporting a Gradient Amplifier X Axis OverVoltage Fault Error: 2244576 Host: WMR2 Process: NSP File: SCP:Grad Line: 2730 The GP is reporting a Gradient Amplifier X Axis OverVoltage Fault.->

2244576 The GP is reporting a Gradient Amplifier Y Axis OverVoltage Fault Error: 2244576 Host: WMR2 Process: NSP File: SCP:Grad Line: 2730 The GP is reporting a Gradient Amplifier Y Axis OverVoltage Fault.->

2244575 The GP is reporting a Gradient Amplifier X Axis OverCurrent Fault Error: 2244575 Host: WMR2 Process: NSP File: SCP:Grad Line: 2683 The GP is reporting a Gradient Amplifier X Axis OverCurrent Fault.->

2244573 The GP reported a Gradient Amplifier % Axis Internal Wiring Fault or Control Undervoltage Error: 2244573 Host: WMR2 Process: NSP File: SCP:Grad Line: 2587 The GP reported a Gradient Amplifier X Axis Internal Wiring Fault or Control Undervoltage.->

2244602 The GP reported multiple Gradient Amplifier X Axis OverVoltage Fault Error: 2244602 Host: WMR2 Process: NSP File: SCP:Grad Line: 2744 The GP reported multiple Gradient Amplifier X Axis OverVoltage Fault

2244575 The GP is reporting a Gradient Amplifier Y Axis OverCurrent Fault Error: 2244575 Host: WMR2 Proc

Applies to

Models
Excite, Signa
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues