Intermittent auto prescan Failure Due to ground connection of incoming power run at the PDU

Symptom

Intermittent auto prescan Failure system would intermittently not scan or stop scan.

Error message

2244589 The GP reported a SGA Power Supply OverTemp Fault 2244589 "The GP reported a SGA Power Supply OverTemp Fault. Wait at least 5 minutes for cabinet to cool. Check fan phase rotation, fan filter, and Gradient Cabinet back door should be closed. Else, check SGA-PS & refer to Service Troubleshooting Manual."

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Area
MR

Full cause & step-by-step resolution